Common faults and solutions of surgical shadowless lamps

News

Surgical lamp

1. The surgical light is not on
Open the top cover and check whether the fuse is blown and whether the power supply voltage is normal. If there are no issues with both, please have them repaired by a professional.
2. Transformer damage
There are generally two reasons for transformer damage, namely power supply voltage issues and overcurrent caused by short circuits.
3. The fuse is often damaged
Check if the shadowless light bulb is configured according to the rated power specified in the manual. If a high-power light bulb is configured, the fuse will be damaged due to its capacity exceeding the rated current of the fuse. Check whether the power supply voltage is normal.
4. Disinfection handle deformed
The disinfection of the shadowless lamp handle can be carried out using high-pressure disinfection, but it should be noted that the handle should not press heavy objects during disinfection, as the fragrance can cause the handle to deform.
5. Turn the shadowless lamp to an angle, and the lamp will not light up
This is mainly because the sensors at both ends of the shadowless lamp suspension rod may have poor contact after use for a period of time, and this situation should be maintained and repaired by a professional.
6. Shadowless lamp displacement
In large surgical shadowless lamps, after use for a period of time, due to the heavy weight of the inner lamp cap, a large amount of friction is required to locate it, which can lead to movement. This can be solved by tightening the upper positioning screw to increase friction.
7. The brightness of the surgical shadowless lamp darkens
The shadowless reflective glass bowl adopts coating technology. General painting techniques can only guarantee a two-year service life, and after two years, the coating may experience problems such as darkening reflections and blistering. So in this situation, you need to replace the reflective bowl.

Surgical lamp.


Post time: Jun-12-2023